DART™ Analysis of Aspirin: Correcting a Misapprehension

FOR IMMEDIATE RELEASE

CONTACT:
Patricia Corkum, Marketing Manager
978-536-2273 • pcorkum@jeol.com • www.jeolusa.com

Introduction

In a recently published comparison1 of the ambient ionization techniques direct analysis in real time (DART™ 2) and DESI3, it was reported that a protonated molecule was not observed for DART, whereas the protonated molecule could be observed for DESI and DAPCI. This is an incorrect observation, resulting from the use of different experimental conditions for DART than were used for the other two techniques. Mass spectra of aspirin measured on a JEOL AccuTOF-DART™ mass spectrometer under the correct operating conditions are shown here. All assignments for the mass spectral peaks were confirmed by exact mass measurements.

Discussion

Because aspirin (acetylsalicylic acid, Figure 1) is an organic acid, it has a relatively low proton affinity. Acidic compounds tend to lose protons rather than gain them. That is, proton transfer from protonated water clusters is not favored.

Figure 1. Aspirin (acetylsalicylic acid, C9H8O4)
Figure 1. Aspirin (acetylsalicylic acid, C9H8O4)

Organic acids tend to deprotonated to form negative ions ([M-H]-). DART analysis in negative-ion mode is the preferred approach for analyzing organic acids. Aspirin is no exception: it readily deprotonates to produce [M-H]- in negative-ion mode (Figure 2).

Figure 2.  Negative-ion DART mass spectrum of aspirin.
Figure 2. Negative-ion DART mass spectrum of aspirin.

If ammonium is present, aspirin will readily form an abundant [M+H]+ and a corresponding [M+NH4]+. This is shown in Figure 3 for aspirin where ammonium is present (from the headspace of a dilute aqueous ammonium hydroxide solution),

Figure 3. Positive-ion DART mass spectrum with ammonium present.
Figure 3. Positive-ion DART mass spectrum with ammonium present.

and in Figure 4 for an over-the-counter analgesic like the tablet described in reference 1. Sampling position was not critical for the DART experiment provided that one does not block the mass spectrometer atmospheric pressure ionization interface. The analgesic tablet was dangled in the DART gas stream, allowing the gas stream to graze the tablet surface to produce the spectra shown in Figure 4.

Conclusion

In reference 1, protonated aspirin was observed in the DESI and DAPCI experiments because ammonium was added. Ammonium was not present for the DART experiment and the protonated molecule was therefore not observed.

Figure 4. DART mass spectra of an over-the-counter analgesic tablet containing aspirin, caffeine, and acetaminophen.
Figure 4. DART mass spectra of an over-the-counter analgesic tablet containing aspirin, caffeine, and acetaminophen.

Reference

  1. “The use of recently described ionisation techniques for the rapid analysis of some common drugs and samples of biological origin”, Jonathan P. Williams, Vibhuti J. Patel, Richard Holland, James H. Scrivens, Rapid Communications in Mass Spectrometry, 20(9), 2006, pp. 1447-1456. http://dx.doi.org/10.1002/rcm.2470
  2. “Versatile New Ion Source for the Analysis of Materials in Open Air under Ambient Conditions” Robert B. Cody, James A. Laramée, and H. Dupont Durst
    Anal. Chem.; 2005; 77(8) pp 2297 - 2302; (Accelerated Article) DOI: 10.1021/ac050162j
  3. DART™ is a trademark of JEOL USA, Inc.
  4. “Mass Spectrometry Sampling Under Ambient Conditions with Desorption Electrospray Ionization”, Zoltán Takáts, Justin M. Wiseman, Bogdan Gologan, R. Graham Cooks, Science, 2004; 306(1595),471-473 http://www.sciencemag.org/cgi/content/abstract/306/5695/471

Related web links


# # #

JEOL USA, Inc.

JEOL is a world leader in electron optical equipment and instrumentation for high-end scientific and industrial research and development. Core product groups include electron microscopes (SEMs and TEMs), instruments for the semiconductor industry (electron beam lithography and a series of defect review and inspection tools), and analytical instruments including mass spectrometers, NMRs and ESRs.

JEOL USA, Inc., is a wholly owned subsidiary of JEOL, Ltd., Japan, was incorporated in the United States in 1962. The company has 13 regional service centers that offer unlimited emergency service and support in the U.S.

For more information about JEOL USA, Inc. or any JEOL products, visit www.jeolusa.com, or call 978-535-5900.


Posted on Thursday, January 31, 2008 (Archive on Friday, January 30, 2009)
Posted by JEOL Webmaster
Return

  

 
  Copyright 2006-2008 JEOL USA, Inc. | Privacy Statement | Terms Of Use
  Designed & Powered by the Swanzey Internet Group