Request Product InfoFind a Local OfficeSearch
 
JEOL
Saturday, May 25, 2013
Login
Register
  PRODUCTS : Medical Equipment : AminoTac™  
Medical Equip. Products
JLC-500/V AminoTac™ Amino Acid Analyzer

JLC-500/VThe AminoTac™ is a next-generation amino acid analyzer. The AminoTac™ attains higher analysis sensitivity of ninhydrin colorimetry, which provides high stability, while featuring a multi-segment tandem column, which provides quick analysis and longer column life. Windows™ based software makes the AminoTac™ easy–to-use and its modular design makes for compact storage and maintenance.

JLC-500/V Key Product Features
Reagent refrigerator Temperature: 5 to 10°C
Number of bottles: 2l × 2,1l × 1
Eluent container Number of bottles: 1l × 5,0.25l × 2
Eluent exchanger Type: Solenoid valve
Number of flow routes:7
Eluent pump Flow sensor feedback and phase-control pulse mixing type
Flow rate constancy under pressure: Within 1%
Maximum operating pressure: 35 MPa
Flow rate range: 0.001 to 1,000 ml/min.
Flow rate accuracy: Within 1 %
Auto sampler Type: Matrix
Accuracy: CV 1% or less (10µl)
Sample amount: 1 to 200µl
Number of settable samples: 96
Temperature: 4 to 10°C
(Thermoelectric cooling device)
Withstand pressure: 35 MPa
Precolumn Packed column
Analytical column Tandem packed column
Column thermostat Type: Thermoelectric heating device
Temperature: 30 to 70°C (variable in 1°C steps)
Reaction bath Type: Cartridge heater
Temperature: 30 to 150°C (variable in 1°C steps)
Detector Type: 3-wavelength (grating)
Measurement wavelength: 440, 570, 690 nm
Light source: Halogen lamp
Controller Programmer: 10 analysis method, regeneration program, wash program
Sefety aids: Pump pressure monitor to protect column, reaction bath temperature, column temperature, eluent & reagent leak detection, reagent level, waste level, etc.
Data processing Detector input: 2 channels (570,440 nm)
Chromatograph data limited run: 3 hours
Calculation: Internal standard, external standard, calibration (0 point-to-5 point fit, linear fit)
Report format: 10 kinds
Print out: Peak name, retention time, peak height, peak area, concentration (mole, gram), amino acid residues, quantity
Reviewing peak: Parameters analyze, mathematical calculation (subtraction, addition, multiplication, and division), manual baseline, calculating derivative, manual peak identification
Dimensions & weight 950(W) × 750(D) × 900(H) mm, 250kg without CRT
Please Note

Not all JEOL Medical Equipment products are available in every country. For specific information and more details about JEOL products available in your area, contact the MED International Business Promotion Group. Thank you.

 
  Copyright 2006-2013 JEOL USA, Inc. | Privacy Statement | Terms Of Use
  Designed & Powered by the Swanzey Internet Group