Identifying structural isomers using retention index - Qualitative analysis of mixed xylenes in lacquer paint using msFineAnalysis iQ
Identifying structural isomers using retention index - Qualitative analysis of mixed xylenes in lacquer paint using msFineAnalysis iQ
MSTips No. 470

Introduction

Typically, qualitative analysis using GC-QMS involves library database (DB) searches using mass spectra obtained by electron ionization (EI) method. However, when qualitative analysis is performed based solely on the similarity to the library spectrum, depending on the compound, multiple significant candidates may be obtained, or the wrong candidate may be selected as the identification result. Therefore, it is effective to use the method in combination with soft ionization (SI) methods such as photoionization (PI) to confirm the molecular ion. In this case, two measurement data, one from the EI method and one from the PI method, are obtained for one sample, making data analysis more complicated, so an integrated qualitative analysis software capable of analyzing the two data quickly and automatically is desirable. To this end, we have developed an integrated qualitative analysis software called "msFineAnalysis iQ" that automatically combines the analysis results of the EI method and the SI method.

However, when the target compound for qualitative analysis contains structural isomers, it may be difficult to distinguish them using only mass spectra. In this case, identification using the retention index (RI) in combination is useful. The retention index is an indicator of the time a compound is retained on a GC column, and each compound has its own unique retention index. Therefore, by comparing the retention index measurement value of an unknown compound with the retention index of a known compound, the compound can be identified.

In this study, a commercially available lacquer paint was measured by pyrolysis GC/MS. Then, integrated qualitative analysis was performed using msFineAnalysis iQ. In this MSTips, we will report an identification example of mixed xylenes (xylol) using retention index.

Attached Files

Other Resources

Walkup NMR
  • See how the Delta NMR software allows users to just "walk up" and start NMR experiments
  • Mass Spec Reference Data
  • View our page of useful molecular references for Mass Spec
  • Tutorials (Mass Spec)
  • Documents on the basics of mass spectrometry
  • Delta NMR software Tutorials
  • Videos on how to use the Delta NMR software
  • No-D NMR
  • Description of No-D NMR and how it can be used to eliminate the need for deuterated solvents
  • Non Uniform Sampling (NUS)
  • Description of how NUS is used to greatly reduce the time needed for running NMR experiments
  • NMR Basics
  • Overview of the Basics of NMR Theory
  • NMR Magnet Destruction
  • See our presentation of the slicing open of a JEOL Delta-GSX 270 MHz NMR Magnet
  • NMR Training
    Basic Operations and System Management for JEOL NMR Users
    Mass Spec Training
    Learn more about spectrometer operation and maintenance, data collection and processing, and advanced MS software operation.
    JEOLink NMR Newsletter
    We publish and send out this NMR newsletter to our customers. They can also be viewed here.
    Mass Media Newsletter
    We publish and send out this Mass Spec newsletter to our customers. They can also be viewed here.
    © Copyright 2026 by JEOL USA, Inc.
    Terms of Use
    |
    Privacy Policy
    |
    Cookie Preferences