Introduction of latest GC/HR-TOFMS system: JMS-T200GC AccuTOF GCx - High Mass Resolution March 5, 2020 AccuTOF™ GC, Application Note, Introduction, Mass Spectrometry (MS) 0 Introduction JEOL has recently announced a 4th generation GC/HR-TOFMS system, the JMS-T200GC AccuTOF GCx, in 2015. The AccuTOF-GCx offers high sensitivity, high mass resolving power, high mass accuracy, and a wide dynamic range in combination with high-speed data acquisition. In this application note, we show the high mass resolution that can be obtained with this latest GC/HR-TOFMS system. Results and Discussion We measured perfluorotributylamine (PFTBA) introduced through the reservoir of the heated volatiles inlet. A resolving power of 10,000 was easily obtained for the m/z 614 peak (Figure 2). Next, we checked the mass separation for a doublet comprised of the peaks corresponding to C6F11+ (PFK) and C7H21O4Si4+ (GC column background). In comparison with the original 1st-generation model, the “AccuTOF GC” (with only a mass resolving power of 5,000), the 4th-generation model “AccuTOF GCX” has over 10,000 mass resolving power (Figure 3). For full details: Attached files often contain the full content of the item you are viewing. Be sure and view any attachments. GCx High Mass Resolution2.pdf 345.97 KB Showing 0 Comment Comments are closed.